Renal stone characterization using high resolution imaging mode on a photon counting detector CT system

Published on Mar 9, 2017in Proceedings of SPIE
· DOI :10.1117/12.2255651
Andrea Ferrero8
Estimated H-index: 8
(Mayo Clinic),
Ralf Gutjahr7
Estimated H-index: 7
(TUM: Technische Universität München)
+ 7 AuthorsCynthia H. McCollough76
Estimated H-index: 76
(Mayo Clinic)
Sources
Abstract
In addition to the standard-resolution (SR) acquisition mode, a high-resolution (HR) mode is available on a research photon-counting-detector (PCD) whole-body CT system. In the HR mode each detector consists of a 2x2 array of 0.225 mm x 0.225 mm subpixel elements. This is in contrast to the SR mode that consists of a 4x4 array of the same subelements, and results in 0.25 mm isotropic resolution at iso-center for the HR mode. In this study, we quantified ex vivo the capabilities of the HR mode to characterize renal stones in terms of morphology and mineral composition. Forty pure stones - 10 uric acid (UA), 10 cystine (CYS), 10 calcium oxalate monohydrate (COM) and 10 apatite (APA) - and 14 mixed stones were placed in a 20 cm water phantom and scanned in HR mode, at radiation dose matched to that of routine dual-energy stone exams. Data from micro CT provided a reference for the quantification of morphology and mineral composition of the mixed stones. The area under the ROC curve was 1.0 for discriminating UA from CYS, 0.89 for CYS vs COM and 0.84 for COM vs APA. The root mean square error (RMSE) of the percent UA in mixed stones was 11.0% with a medium-sharp kernel and 15.6% with the sharpest kernel. The HR showed qualitatively accurate characterization of stone morphology relative to micro CT.
📖 Papers frequently viewed together
5 Citations
9 Citations
14 Citations
References19
Newest
#1Shuai Leng (Mayo Clinic)H-Index: 53
#2Alice Huang (Mayo Clinic)H-Index: 1
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 6 authors...
OBJECTIVE. The purpose of this study was to assess the feasibility of using dual-energy CT to accurately quantify uric acid and non–uric acid components in urinary stones of mixed composition. MATERIALS AND METHODS. A total of 24 urinary stones were analyzed with micro CT to serve as the reference standard for uric acid and non–uric acid composition. These stones were placed in water phantoms to simulate body attenuation of slim to obese adults and scanned with a third-generation dual-source CT ...
18 CitationsSource
#1Shuai Leng (Mayo Clinic)H-Index: 53
#2Zhicong Yu (Mayo Clinic)H-Index: 11
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 12 authors...
A high-resolution (HR) data collection mode has been introduced to a whole-body, research photon-counting-detector CT system installed in our laboratory. In this mode, 64 rows of 0.45 mm x 0.45 mm detector pixels were used, which corresponded to a pixel size of 0.25 mm x 0.25 mm at the iso-center. Spatial resolution of this HR mode was quantified by measuring the MTF from a scan of a 50 micron wire phantom. An anthropomorphic lung phantom, cadaveric swine lung, temporal bone and heart specimens ...
14 CitationsSource
#1Zhicong Yu (Mayo Clinic)H-Index: 11
#2Shuai Leng (Mayo Clinic)H-Index: 53
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 11 authors...
This study evaluated the conventional imaging performance of a research whole-body photon-counting CT system and investigated its feasibility for imaging using clinically realistic levels of x-ray photon flux. This research system was built on the platform of a 2nd generation dual-source CT system: one source coupled to an energy integrating detector (EID) and the other coupled to a photon-counting detector (PCD). Phantom studies were conducted to measure CT number accuracy and uniformity for wa...
140 CitationsSource
#1Xinhui Duan (Mayo Clinic)H-Index: 18
#2Zhoubo Li (Mayo Clinic)H-Index: 16
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 7 authors...
OBJECTIVE. The purpose of this phantom study was to determine the utility of a third-generation dual-source CT scanner with increased dual-energy spectral separation in differentiating urinary stone composition. MATERIALS AND METHODS. Eighty-seven urinary stones from humans were scanned in 35-, 40-, 45-, and 50-cm wide anthropomorphic phantoms with a third-generation dual-source scanner (system A) with a high-energy beam of 150 kV plus 0.6-mm tin filtration (Sn). The low-energy data were acquire...
27 CitationsSource
#1Shuai LengH-Index: 53
#2Felix E. DiehnH-Index: 25
Last. Cynthia H. McColloughH-Index: 76
view all 7 authors...
BACKGROUND AND PURPOSE: Radiation dose in temporal bone CT imaging can be high due to the requirement of high spatial resolution. In this study, we assessed whether CT imaging of the temporal bone by using an ultra-high-resolution scan mode combined with iterative reconstruction provides higher spatial resolution and lower image noise than a z-axis ultra-high-resolution mode. MATERIALS AND METHODS: Patients with baseline temporal bone CT scans acquired by using a z-axis ultra-high-resolution pro...
21 CitationsSource
#1Katsuyuki Taguchi (JHUSOM: Johns Hopkins University School of Medicine)H-Index: 26
#2Jan S. IwanczykH-Index: 28
Photon counting detectors (PCDs) with energy discrimination capabilities have been developed for medical x-ray computed tomography (CT) and x-ray (XR) imaging. Using detection mechanisms that are completely different from the current energy integrating detectors and measuring the material information of the object to be imaged, these PCDs have the potential not only to improve the current CT and XR images, such as dose reduction, but also to open revolutionary novel applications such as molecula...
522 CitationsSource
#1Cynthia H. McCollough (Mayo Clinic)H-Index: 76
#2Shuai Leng (Mayo Clinic)H-Index: 53
Last. Thomas Flohr (Siemens)H-Index: 86
view all 9 authors...
Purpose: To assess the z-axis resolution improvement and dose reduction potential achieved using a z-axis deconvolution technique with iterative reconstruction (IR) relative to filtered backprojection (FBP) images created with the use of a z-axis comb filter. Methods: Each of three phantoms were scanned with two different acquisition modes: (1) an ultrahigh resolution (UHR) scan mode that uses a comb filter in the fan angle direction to increase in-plane spatial resolution and (2) a z-axis ultra...
22 CitationsSource
#1Polad M Shikhaliev (LSU: Louisiana State University)H-Index: 9
Photon counting spectral computed tomography (PCSCT) provides material selective CT imaging at a single CT scan and fixed tube voltage. The PCSCT data are acquired in several energy ranges (bins) arranged over the x-ray spectrum. The quasi-monoenergetic CT images are acquired in these energy bins and are used for material decomposition. The PCSCT exhibits inherent limitations when material decomposition is performed using energy bins. For effective material decomposition, the energy bins used fo...
54 CitationsSource
#1Ravi K. Kaza (UM: University of Michigan)H-Index: 21
#2Joel F. Platt (UM: University of Michigan)H-Index: 53
Last. Ashish P. WasnikH-Index: 15
view all 6 authors...
Proper understanding of dual-energy CT image data processing methods and appropriate use of material-specific and quantitative information allow improved patient care and further refinement in genitourinary applications of dual-energy CT.
141 CitationsSource
#1Mingliang Qu (Mayo Clinic)H-Index: 11
#2Juan Carlos Ramirez-Giraldo (Mayo Clinic)H-Index: 16
Last. Cynthia H. McCollough (Mayo Clinic)H-Index: 76
view all 7 authors...
OBJECTIVE. The purpose of this study was to determine the ex vivo ability of dual-energy dual-source CT (DSCT) with additional tin filtration to differentiate among five groups of human renal stone types. MATERIALS AND METHODS. Forty-three renal stones of 10 types were categorized into five primary groups on the basis of effective atomic numbers, which were calculated as the weighted average of the atomic numbers of constituent atoms. Stones were embedded in porcine kidneys and placed in a 35-cm...
104 CitationsSource
Cited By4
Newest
Photon counting x-ray detectors (PCDs) with spectral capabilities have the potential to revolutionize computed tomography (CT) for medical imaging. The ideal PCD provides accurate energy information for each incident x-ray, and at high spatial resolution. This information enables material-specific imaging, enhanced radiation dose efficiency, and improved spatial resolution in CT images. In practice, PCDs are affected by non-idealities, including limited energy resolution, pulse pileup, and cross...
4 CitationsSource
#1Megan C. Jacobsen (University of Texas MD Anderson Cancer Center)H-Index: 7
#2Sara L. Thrower (University of Texas MD Anderson Cancer Center)H-Index: 2
Last. Rick R. Layman (University of Texas MD Anderson Cancer Center)H-Index: 9
view all 10 authors...
Computed tomography (CT) technology has rapidly evolved since its introduction in the 1970s. It is a highly important diagnostic tool for clinicians as demonstrated by the significant increase in utilization over several decades. However, much of the effort to develop and advance CT applications has been focused on improving visual sensitivity and reducing radiation dose. In comparison to these areas, improvements in quantitative CT have lagged behind. While this could be a consequence of the te...
2 CitationsSource
#1Muhammad U. Ghani (OU: University of Oklahoma)H-Index: 7
#2Yuhua Li (OU: University of Oklahoma)H-Index: 12
Last. Hong Liu (OU: University of Oklahoma)H-Index: 33
view all 4 authors...
Abstract The recent advancements in the photon counting detection have created a significant growing research interest in the x-ray imaging. It is essential to objectively understand the image quality parameters of a photon counting detector before developing imaging applications. In this work, we have assessed the imaging quality of a cadmium telluride (CdTe) based PCD in projection imaging mode. The detector is 70.4 mm × 6.6 mm dimensions. The detector has a pixel array of 64 × 4 with a pixel ...
1 CitationsSource
#1Bhavik N. Patel (Stanford University)H-Index: 20
#2Daniele Marin (Duke University)H-Index: 39
: Dual-energy computed tomography (DECT) offers several advantages over conventional single-energy CT. These advantages include improved image quality, beam hardening correction, and metal artifact reduction. Additionally, DECT allows derivation of quantitative information through material decomposition analysis. Although newer third-generation rapid-kilovolt switching and dual-source DECT scanners have significantly improved in image quality and workflow compared with initial iterations and ear...
9 CitationsSource