High-sensitivity intravascular photoacoustic imaging of lipid–laden plaque with a collinear catheter design

Volume: 6, Issue: 1
Published: Apr 28, 2016
Abstract
A highly sensitive catheter probe is critical to catheter-based intravascular photoacoustic imaging. Here, we present a photoacoustic catheter probe design on the basis of collinear alignment of the incident optical wave and the photoacoustically generated sound wave within a miniature catheter housing for the first time. Such collinear catheter design with an outer diameter of 1.6 mm provided highly efficient overlap between optical and...