Abstract
We evaluated independently retrospective half scan and multisector mode manufacturer's protocols and compared them with modified acquisition protocols to determine optimal imaging parameters for cardiac scanning.Data were acquired using two fabricated gated moving phantoms. In half scan mode, the manufacturer's recommended pitch values were compared with adjacent values at different motion rates. In multisector mode, the manufacturer's protocols...
Paper Details
Title
Evaluation of Retrospective Multisector and Half Scan ECG-Gated Multidetector Cardiac CT Protocols with Moving Phantoms
Published Date
Sep 1, 2002
Volume
26
Issue
5
Pages
768 - 776
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.