Image quality evaluation of flat panel and image intensifier digital magnification in x‐ray fluoroscopy

Volume: 29, Issue: 7, Pages: 1611 - 1621
Published: Jun 24, 2002
Abstract
Interventional devices used in radiology often have dimensions on the order of a pixel, and radiologists resort to image magnification to better visualize such small devices. Traditional image intensifier (II) systems use analog magnification with x-ray exposure inversely proportional to the area of field of view (FOV) so as to maintain light output for the camera. Analog magnification is impossible with flat panel (FP) detectors, and images...
Paper Details
Title
Image quality evaluation of flat panel and image intensifier digital magnification in x‐ray fluoroscopy
Published Date
Jun 24, 2002
Volume
29
Issue
7
Pages
1611 - 1621
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.