Scanning-beam digital x-ray (SBDX) system for cardiac angiography

Published on May 28, 1999
路 DOI :10.1117/12.349499
Edward G. Solomon8
Estimated H-index: 8
,
Brian P. Wilfley8
Estimated H-index: 8
+ 2 AuthorsJoseph A. Heanue9
Estimated H-index: 9
Sources
Abstract
An advanced Scanning-Beam Digital X-ray (SBDX) system for cardiac angiography has been constructed. The 15-kW source operates at 70 - 120 kVp and has an electron beam that is electromagnetically scanned across a 23-cm X 23-cm transmission target. The target is directly liquid cooled for continuous full-power operation and is located behind a focused source collimator. The collimator is a rectangular grid of 100 X 100 apertures whose axes are aligned with the center of the detector array. X-ray beam divergence through the collimator apertures is matched to the 5.4-cm X 5.4 cm detector, which is 150 cm from the source. The detector is a 48 X 48 element CdZnTe direct-conversion photon-counting detector. A narrow x-ray beam scans the full field of view at up to 30 frames per second. A custom digital processor simultaneously reconstructs sixteen 1,0002 pixel tomographic images in real time. The slices are spaced 1.2 cm apart and cover the entire cardiac anatomy. The small detector area and large patient-detector distance result in negligible detected x-ray scatter. Image signal-to-noise ratio is calculated to be equal to conventional fluoroscopic systems at only 12% of the patient exposure and 25% of the staff exposure. Exposure reduction is achieved by elimination of detected scatter, elimination of the anti-scatter grid, increased detector DQE, and increased patient entrance area.漏 (1999) COPYRIGHT SPIE--The International Society for Optical Engineering. Downloading of the abstract is permitted for personal use only.
馃摉 Papers frequently viewed together
References5
Newest
#1Marian C. Limacher (American College of Cardiology)H-Index: 2
#2Pamela S. DouglasH-Index: 123
Last. Mary Norine WalshH-Index: 53
view all 10 authors...
Source
#1Michael S. Van Lysel (UW: University of Wisconsin-Madison)H-Index: 11
#2Edward G. SolomonH-Index: 8
Last. Michael A. Speidel (UW: University of Wisconsin-Madison)H-Index: 11
view all 5 authors...
A prototype scanning-beam digital x-ray (SBDX) system for cardiac fluoroscopy has been constructed. The unique geometry and absence of detected x-ray scatter in the SBDX image promises to provide image quality equivalent to a conventional image-intensifier-based fluoroscopic system at substantially reduced x-ray exposure to patient and staff. In order to measure the SBDX exposure advantage, a contrast- detail study was performed comparing SBDX and a conventional cardiac fluoroscopic system. Low-...
Source
#1Edward G. SolomonH-Index: 8
#2Michael S. Van Lysel (UW: University of Wisconsin-Madison)H-Index: 11
Last. Brian SkillicornH-Index: 1
view all 5 authors...
A prototype scanning-beam digital x-ray system for cardiac fluoroscopy has been constructed. Source-to-detector distance is 94 cm with the subject positioned near the source. The 4-kW source operates at 70-110 kVp and has an electromagnetically-scanned 25-cm-diameter transmission target. The target is at ground potential and is directly liquid cooled for continuous full-power operation. The source collimator has 22,000 holes whose axes are aligned with the center of the detector array. Beam dive...
Source
#1Hiroshi Fujita (Gidai: Gifu University)H-Index: 64
#2Du-Yih TsaiH-Index: 13
Last. A. Ohtsuka (Gidai: Gifu University)H-Index: 1
view all 7 authors...
The authors developed a simple method for determining the presampling modulation transfer function (MTF). which includes the unsharpness of the detector and the effect of the sampling aperture, in digital radiographic (DR) systems. With this method, the presampling MTF is determined by the Fourier transform of a 'finely sampled' line spread function (LSF) obtained with a slightly angulated slit in a single exposure. Since the effective sampling distance becomes much smaller than the original sam...
Source
Source
Cited By36
Newest
#1Derrek Spronk (UNC: University of North Carolina at Chapel Hill)H-Index: 7
#2Yueting Luo (UNC: University of North Carolina at Chapel Hill)H-Index: 1
Last. Otto Zhou (UNC: University of North Carolina at Chapel Hill)H-Index: 69
view all 6 authors...
PURPOSE Stationary computed tomography (s-CT) conceptually offers several advantages over existing rotating-gantry based CT. Over the last 40 years, s-CT has been investigated using different technological approaches. We are developing a s-CT system specifically for head/brain imaging using carbon nanotube (CNT) based field emission x-ray source array technology. The non-circular geometry requires different assessment approaches as compared to circular geometries. The purpose of the present stud...
Source
#1Kang An (Chongqing University)H-Index: 1
#2Yifan Yin (Chongqing University)H-Index: 1
Last. Rifeng Zhou (Chongqing University)H-Index: 1
view all 7 authors...
We developed a new prototype micro-multifocus X-ray source based on electron beam scanning (MMFEBS) for multi-view stationary Micro Computed Tomography (multi-view stationary Micro-CT). This micro-multifocus X-ray source includes an electron gun, electron beam deflection coils, a dynamic condensor, a flat panel transmission target and dynamic vacuum system. The electron beam is controlled by the deflection coils and sweeps back and forth on the flat panel transmission target. It hits a series of...
Source
#1Khanlian Chung (Heidelberg University)H-Index: 4
#2Lothar R. Schad (Heidelberg University)H-Index: 66
Last. Frank G. Z枚llner (Heidelberg University)H-Index: 21
view all 3 authors...
Purpose Cone beam computed tomography (CBCT) systems offer physicians crucial 3D and 2D imaging capabilities during interventions. However, certain medical applications only require very specific information from the CBCTs (e.g., determination of the position of high-contrast objects). In diagnostics, tomosynthesis techniques can be used in these cases to minimize dose exposure. Therefore, integrating such techniques on CBCT systems could also be beneficial for interventions. In this paper, we i...
Source
#1Scott S. Hsieh (UCLA: University of California, Los Angeles)H-Index: 13
#2Lydia W. Ng (SC: University of Southern California)H-Index: 2
Purpose Fluoroscopy has been a tool of choice for monitoring treatments or interventions because of its extremely fast imaging times. However, the contrast obtained in fluoroscopy may be insufficient for certain clinical applications. In stereotactic ablative radiation therapy of the lung, fluoroscopy often lacks sufficient contrast for gating treatment. The purpose of this work is to describe and assess a real-time tomosynthesis design that can produce sufficient contrast for guidance of lung t...
Source
#1Bruno De Man (GE: General Electric)H-Index: 20
#2Jorge Uribe (GE: General Electric)H-Index: 6
Last. Norbert J. Pelc (Stanford University)H-Index: 78
view all 16 authors...
Purpose: This paper presents an overview of multisource inverse-geometry computed tomography (IGCT) as well as the development of a gantry-based research prototype system. The development of the distributed x-ray source is covered in a companion paper [V. B. Neculaes et al., 鈥淢ultisource inverse-geometry CT. Part II. X-ray source design and prototype,鈥 Med. Phys. 43, 4617鈥4627 (2016)]. While progress updates of this development have been presented at conferences and in journal papers, this paper...
Source
#1Meng Wu (Stanford University)H-Index: 6
#2Sungwon Yoon (Varian Medical Systems)H-Index: 3
Last. Rebecca Fahrig (Stanford University)H-Index: 33
view all 6 authors...
Digital tomosynthesis is a three-dimensional imaging technique with a lower radiation dose than computed tomography (CT). Due to the missing data in tomosynthesis systems, out-of-plane structures in the depth direction cannot be completely removed by the reconstruction algorithms. In this work, we analyzed the impulse responses of common tomosynthesis systems on a plane-to-plane basis and proposed a fast and accurate convolution-based blur-and-add (BAA) model to simulate the backprojected images...
Source
#1Meng Wu (Stanford University)H-Index: 6
#2Rebecca Fahrig (Stanford University)H-Index: 33
Purpose: The scanning beam digital x-ray system (SBDX) is an inverse geometry fluoroscopic system with high dose efficiency and the ability to perform continuous real-time tomosynthesis in multiple planes. This system could be used for image guidance during lung nodule biopsy. However, the reconstructed images suffer from strong out-of-plane artifact due to the small tomographic angle of the system. Methods: The authors propose an out-of-plane artifact subtraction tomosynthesis (OPAST) algorithm...
Source
#1Prasheel Lillaney (UCSF: University of California, San Francisco)H-Index: 10
#2Norbert J. Pelc (Stanford University)H-Index: 78
Last. Rebecca Fahrig (Stanford University)H-Index: 33
view all 3 authors...
Minimally invasive procedures using image guidance have become a popular alternative to their surgical counterparts because they offer reduced patient risk and as a result reduced morbidity. X-ray fluoroscopy can be used to provide real-time image guidance for interventional procedures. Current state-of-the-art fluoroscopy systems consist of a high-power rotating anode x-ray tube and a digital flat-panel detector. However, the lack of three-dimensional visualization has motivated the development...
Source
#1G Nelson (Stanford University)H-Index: 6
#2Sungwon Yoon (Varian Medical Systems)H-Index: 3
Last. Rebecca Fahrig (Stanford University)H-Index: 33
view all 5 authors...
Purpose: An improved method of image guidance for lung tumor biopsies could help reduce the high rate of false negatives. The aim of this work is to optimize the geometry of the scanning-beam digital tomography system (SBDX) for providing real-time 3D tomographic reconstructions for target verification. The unique geometry of the system requires trade-offs between patient dose, imaging field of view (FOV), and tomographic angle. Methods: Tomosynthetic angle as a function of tumor-to-detector dis...
Source
#1Scott S. Hsieh (Stanford University)H-Index: 13
#2Joseph A. HeanueH-Index: 9
Last. Norbert J. Pelc (Stanford University)H-Index: 78
view all 7 authors...
Purpose: Inverse geometry computed tomography (IGCT) has been proposed as a new system architecture that combines a small detector with a large, distributed source. This geometry can suppress cone-beam artifacts, reduce scatter, and increase dose efficiency. However, the temporal resolution of IGCT is still limited by the gantry rotation time. Large reductions in rotation time are in turn difficult due to the large source array and associated power electronics. We examine the feasibility of usin...
Source
This website uses cookies.
We use cookies to improve your online experience. By continuing to use our website we assume you agree to the placement of these cookies.
To learn more, you can find in our Privacy Policy.