<title>Low-exposure scanning-beam x-ray fluoroscopy system</title>

Published: Apr 11, 1996
Abstract
A prototype scanning-beam digital x-ray system for cardiac fluoroscopy has been constructed. Source-to-detector distance is 94 cm with the subject positioned near the source. The 4-kW source operates at 70-110 kVp and has an electromagnetically-scanned 25-cm-diameter transmission target. The target is at ground potential and is directly liquid cooled for continuous full-power operation. The source collimator has 22,000 holes whose axes are...
Paper Details
Title
<title>Low-exposure scanning-beam x-ray fluoroscopy system</title>
Published Date
Apr 11, 1996
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.