Summary.  The Platelet Function Analyzer (PFA-100®) was used to measure platelet function in paediatric patients with destructive versus underproduction thrombocytopenia. Closure time (CT) and total volume (TV) measurements with standard 150 μm apertures discriminated between patients with similar platelet counts from 30 to 150 × 109/l. However, at platelet counts < 30 × 109/l, a 100-μm aperture (experimental) gave the best assessment of...
Paper Fields
Paper Details
Title
Assessment of thrombocytopenic disorders using the Platelet Function Analyzer (PFA-100).
Published Date
Jun 1, 2002
Volume
117
Issue
4
Pages
961 - 964