The study was aimed to check the radiotherapy treatment accuracy and definition of action levels during implementation of in vivo dosimetry as a part of quality assurance program. The calibration and correction factors for in vivo entrance dose measurements for six n-type Isorad semiconductor diodes were determined as recommended by the European Society for Radiotherapy and Oncology Booklet No. 5. The patients for in vivo measurements have been...