Metal Artifact Reduction With MAVRIC SL at 3-T MRI in Patients With Hip Arthroplasty

Volume: 204, Issue: 1, Pages: 140 - 147
Published: Jan 1, 2015
Abstract
The objective of our study was to compare the multiacquisition variable-resonance image combination selective (MAVRIC SL) sequence with the 2D fast spin-echo (FSE) sequence for metal artifact reduction on 3-T MRI in patients with hip arthroplasty (HA).Matched 2D FSE and MAVRIC SL images of 21 hips (19 patients with HA) were included in the study group. Paired image sets, composed of 13 coronal and 12 axial slices (total, 25 image sets), of the...
Paper Details
Title
Metal Artifact Reduction With MAVRIC SL at 3-T MRI in Patients With Hip Arthroplasty
Published Date
Jan 1, 2015
Volume
204
Issue
1
Pages
140 - 147
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.