Comparison of measured and Monte Carlo calculated dose distributions from the NRC linac
Abstract
We have benchmarked photon beam simulations with the EGS4 user code BEAM [Rogers et al., Med. Phys. 22, 503-524 (1995)] by comparing calculated and measured relative ionization distributions in water from the 10 and 20 MV photon beams of the NRC linac. Unlike previous calculations, the incident electron energy is known independently to 1%, the entire extra-focal radiation is simulated, and electron contamination is accounted for. The full Monte...
Paper Details
Title
Comparison of measured and Monte Carlo calculated dose distributions from the NRC linac
Published Date
Oct 1, 2000
Journal
Volume
27
Issue
10
Pages
2256 - 2266
Citation AnalysisPro