Combined MR-assisted motion and partial volume effects corrections – impact on PET data quantification

Published on Jul 29, 2014in EJNMMI Physics2.574
· DOI :10.1186/2197-7364-1-S1-A38
Ciprian Catana40
Estimated H-index: 40
(Harvard University),
Daniel B. Chonde17
Estimated H-index: 17
(MIT: Massachusetts Institute of Technology)
+ 4 AuthorsJoshua L. Roffman41
Estimated H-index: 41
(Harvard University)
Sources
Abstract
Our goal in this study was to characterize the combined effect of MR-assisted motion correction (MC) and partial volume effects correction (PVEC) on the estimation of [11C]NNC112 binding potential (BP) in healthy volunteers. 29 subjects were scanned on the Siemens 3T MR-BrainPET scanner prototype. Emission data were acquired in list mode format for 90-minutes following the i.v. administration of ~8 mCi of [11C]NNC112. The head attenuation map was obtained from the MPRAGE data using an atlas-based method. Head motion estimates were derived from the MR data and used to correct the PET data in LOR space before image reconstruction [1]. PVEC was applied to the motion corrected data using the region-based voxel-wise (RBV) method [2] and regions of interest (ROIs) defined from the MPRAGE images using FreeSurfer and the measured point-spread function [3]. BPnd for each of the ROIs was estimated in PMOD using the simplified reference tissue kinetic (SRTM) model and the cerebellum as a reference tissue. Maximum translations of up to 9 mm and rotations of up to 12 degrees have been observed in this group of subjects (Figure ​(Figure1).1). Less variability in the tissue time activity curves (TACs) was noted after MC (the curves before and after MC for a representative subject are shown in Figure ​Figure2).2). The percentage changes in BPnd after MC and PVEC revealed both under- and overestimation in the ROIs analyzed (Figure ​(Figure3).3). The cumulative effect exceeded 100% for some of the structures analyzed. Figure 1 Maximum translations (left) and rotations (right) measured in 29 healthy volunteers scanned for 90 minutes Figure 2 TACs in the left and right caudate and cerebellar cortices before (left) and after (right) MR-assisted motion correction Figure 3 Percent change in NNC112 BPnd after motion and partial volume effects corrections Significant motion and PVE were observed in all the subjects, biasing the PET estimates. The combined effect is difficult to predict, depending on the size and location of the structure of interest and patient compliance. Without addressing these issues, the value of the BPnd’s derived from these data is questionable.
📖 Papers frequently viewed together
11 Authors (Sune H. Keller, ..., Adam E. Hansen)
3 Citations
1 Citations
203 Citations
References3
Newest
#1Spencer L. BowenH-Index: 11
#3Christian Michel (Siemens)H-Index: 20
Last. Ciprian CatanaH-Index: 40
view all 5 authors...
Kinetic parameters estimated from dynamic 18F-fluorodeoxyglucose (18F-FDG) PET acquisitions have been used frequently to assess brain function in humans. Neglecting partial volume correction (PVC) for a dynamic series has been shown to produce significant bias in model estimates. Accurate PVC requires a space-variant model describing the reconstructed image spatial point spread function (PSF) that accounts for resolution limitations, including non-uniformities across the field of view due to the...
17 CitationsSource
#1Benjamin A. Thomas (UCL: University College London)H-Index: 10
#2Kjell Erlandsson (UCL: University College London)H-Index: 27
Last. Brian Hutton (UCL: University College London)H-Index: 66
view all 7 authors...
Purpose Alzheimer’s disease (AD) is the most common form of dementia. Clinically, it is characterized by progressive cognitive and functional impairment with structural hallmarks of cortical atrophy and ventricular expansion. Amyloid plaque aggregation is also known to occur in AD subjects. In-vivo imaging of amyloid plaques is now possible with positron emission tomography (PET) radioligands. PET imaging suffers from a degrading phenomenon known as the partial volume effect (PVE). The quantitat...
200 CitationsSource
#1Ciprian Catana (Harvard University)H-Index: 40
#2Thomas BennerH-Index: 49
Last. A. Gregory SorensenH-Index: 82
view all 10 authors...
Head motion is difficult to avoid in long PET studies, degrading the image quality and offsetting the benefit of using a highresolution scanner. As a potential solution in an integrated MRPET scanner, the simultaneously acquired MRI data can be used for motion tracking. In this work, a novel algorithm for data processing and rigid-body motion correction (MC) for the MRIcompatible BrainPET prototype scanner is described, and proof-of-principle phantom and human studies are presented. Methods: To ...
151 CitationsSource
Cited By1
Newest
#1Liliana Caldeira (Forschungszentrum Jülich)H-Index: 7
#2Nuno Andre da Silva (Forschungszentrum Jülich)H-Index: 5
Last. N. Jon Shah (Forschungszentrum Jülich)H-Index: 58
view all 5 authors...
Dynamic PET provides temporal information about the tracer uptake. However, each PET frame has usually low statistics, resulting in noisy images. Furthermore, PET images suffer from partial volume effects. The goal of this study is to understand the effects of prior regularisation on dynamic PET data and subsequent anatomical partial volume correction. The Median Root Prior (MRP) regularisation method was used in this work during reconstruction. The quantification and noise in image-domain and t...
2 CitationsSource