Dosimetry tools and techniques for IMRT
Abstract
Intensity modulated radiation therapy (IMRT) poses a number of challenges for properly measuring commissioning data and quality assurance (QA) radiation dose distributions. This report provides a comprehensive overview of how dosimeters, phantoms, and dose distribution analysis techniques should be used to support the commissioning and quality assurance requirements of an IMRT program. The proper applications of each dosimeter are described...
Paper Details
Title
Dosimetry tools and techniques for IMRT
Published Date
Feb 16, 2011
Journal
Volume
38
Issue
3
Pages
1313 - 1338