Six years experience in intracoronary brachytherapy procedures: patient doses from fluoroscopy
Abstract
Typical patient dose levels during intracoronary brachytherapy (ICB) procedures using beta sources were determined across a sample of 221 treatments. Dose-area product values, fluoroscopy time and number of frames per procedure, with median values of 62 Gy cm2, 17.0 min and 1493 images, respectively, resulted in a 20% to 50% increase in the values measured for percutaneous transluminal coronary angioplasty procedures in the same medical centre...
Paper Details
Title
Six years experience in intracoronary brachytherapy procedures: patient doses from fluoroscopy
Published Date
Sep 1, 2006
Journal
Volume
79
Issue
945
Pages
730 - 733