A multi-source inverse-geometry CT system: initial results with an 8 spot x-ray source array

Volume: 59, Issue: 5, Pages: 1189 - 1202
Published: Feb 20, 2014
Abstract
We present initial experimental results of a rotating-gantry multi-source inverse-geometry CT (MS-IGCT) system. The MS-IGCT system was built with a single module of 2 × 4 x-ray sources and a 2D detector array. It produced a 75 mm in-plane field-of-view (FOV) with 160 mm axial coverage in a single gantry rotation. To evaluate system performance, a 2.5 inch diameter uniform PMMA cylinder phantom, a 200 µm diameter tungsten wire, and a euthanized...