Comprehensive quality assurance phantom for the small animal radiation research platform (SARRP)

Volume: 31, Issue: 5, Pages: 529 - 535
Published: Jul 1, 2015
Abstract
To develop and test the suitability and performance of a comprehensive quality assurance (QA) phantom for the Small Animal Radiation Research Platform (SARRP).A QA phantom was developed for carrying out daily, monthly and annual QA tasks including: imaging, dosimetry and treatment planning system (TPS) performance evaluation of the SARRP. The QA phantom consists of 15 (60 × 60 × 5 mm(3)) kV-energy tissue equivalent solid water slabs. The phantom...
Paper Details
Title
Comprehensive quality assurance phantom for the small animal radiation research platform (SARRP)
Published Date
Jul 1, 2015
Volume
31
Issue
5
Pages
529 - 535
Citation AnalysisPro
  • Scinapse’s Top 10 Citation Journals & Affiliations graph reveals the quality and authenticity of citations received by a paper.
  • Discover whether citations have been inflated due to self-citations, or if citations include institutional bias.